This product is used to measure the quality of breathing air delivered by a high-pressure system (e.g. compressor or compressed air cylinder). Dräger tubes ® or Dräger Impactor are used for measurement. The values can be measured both simultaneously or individually. All components of the Aerotest Simultan HP FN are arranged in a carrying case ready for use. Delivery includes: 1 pressure reducer with manual connection 1 measuring device for 4 tubes 1 timer 1 break-off device for tubes 1 bubble test hose 1 package of 5 sinter filters 1 instructions for use 1 transport case (black) Technical data: Carrying case, LxWxH in mm: 350x300x85 Weight incl. contents: approx. 3 kg Pressure reducer: Input: male thread M25x2-300 bar (manual connection) Admission pressure max.: 300 bar Pressure gauge: 0 to 400 bar Output: barbed nipple 9 mm for measuring device Relief valve: set to 10 ±2 bar Ambient conditions: 15 °C to 25 °C, 1013 mbar Compressed air supply connection: barbed nipple 9 mm Admission pressure (pre-set upon delivery): 8.5 bar Flow CO and CO2 tube: 0.2 l/min Impactor and H2O tube: 4 l/min